Do you have a question that is not listed in this section? Please contact our Helpdesk!
+
Frequently Asked Questions Demcon convergence SDI/MFI
Frequently Asked Questions Demcon convergence SDI/MFI
1. Maintenance related questions
1.1. In what specific cases should the user clean the SDI Inspector with a cleaning agent?
There are three triggers that can indicate fouling of the system.
1. Too many Flow errors.
2. MFI is above 1.5 with clean water.
3. Visible fouling in the collected water.
1.2. How often cleaning without cleaning agent is recommended to be performed?
Each time fouled water has been used it is advised to run this cycle with clean water when the tests have finished.
1.3. How do we know if the cleaning was effectively performed?
If the user was triggered to perform a cleaning cycle with agent, they should check the following points:
1. Are the flow errors gone?
2. Has the MFI value dropped to normal values with clean water. (between 0.5-1.5)?
3. Do they still see visible fouling come out of the instrument?
1.4. How should I prepare my system for longer storage?
1. At first a cleaning with agent must be performed. When finished, remove the tubing from the IN-inlet.
2. Start a Cleaning cycle without agent, when no water comes out the outlet, abort the cycle.
3. Remove the tubing from the IN/OUT ports and place the RED/Black stop plugs instead.
4. Remove the Fuse and store it in a safe place, close to the system.
1.5. Why should I remove the fuse when the instrument is not in use?
The battery discharges more quickly when the fuse is in the instrument.
1.6. What kind of clean water should be used when flushing the system?
Tap water can be used, but preferably filtered distilled water with a lower conductivity should be used.
2. Instrument related questions
2.1. How can I calibrate my SDI?
You can use a reference liquid to be used use as a Calibrator. We suggest distilled water and repeat 5 SDI tests (with each time a new filter) and collect the values. This will give you a good reference for future tests.
2.2. What type of tubing is used for the IN/OUT lines?
We use PE (PolyEthyleen) (6mm x 4mm). It should be hard material, to be used in Push-ins.
2.3. How long can the tubing to the Feed can be extended.
Not more than 2 meters. The inner diameter of the tube is 4mm which needs some time to fill. It should be made sure that during the deaeration process (30 s long) the water can reach the membrane filter.
2.4. Can I connect my charger during the actual test?
Yes, this is no problem. It will not affect the test that is running.
3. Test related questions
3.1. What sample volume is needed for a single test?
One test usually requires 2,0 L – 2,5 L of feed solution.
3.2. Can I use a filter after a quick test abort?
Depending on at which stage the test was aborted. If it was in the very beginning, you can still use it. If the test has already been running for longer, you need to use a new filter. Otherwise the experiments will not be accurate since fouling has already occurred on the filter.
3.3. What information is shown on the Graph during the test?
During the test next Flow Graphs are shown
1. Green: Pressure
2. Blue: Flow
3. Red: Measured Volume
Further we see the following screen indicators:
4. Time: Elapsed time (Nominal test time is 20 minutes)
5. Sigma V: Total volume is calculated and in the final results
6. Temp: Actual temperature of the Feed water
7. MFI: Actual calculated MFI value
8. SDIx: Calculated values
3.4. What should I do when I have Pressure errors?
1. Check if the tubing is correctly positioned in the Feed
2. Check if your SDI filter is correctly positioned.
3. Check if the tubing underneath the filter is positioned correctly.
4. Replace the filter
5. Contact our Helpdesk for help.
3.5. What problems can should I do when I have Flow errors?
1. Check if no air is aspirated from the flow
2. Repeat the test with the same filter
3. Check during this test if water is coming out from the Output.
4. If still the same, replace the filter and repeat the test
5. If there are still flow errors, clean the system.
6. Contact our Helpdesk for help.
4. Results related questions
4.1. What conclusion can I draw from my SDI results?
Next table give the following indications for particulate fouling:
SDI < 1.5 Very Clean
SDI 1.5 – 3.0 Good and acceptable
SDI 3.0 – 5.0 Fouling indication
SDI > 5.0 Excessive fouling
4.2. What MFI I would expect with clean tap water of distilled water?
If the SDI unit has been maintained properly, value between 0.5 and 1.5 are values to be expected.
4.3. How many results can be stored in the SDI?
The system can store up to a 1000 measurements
+
When do I get an answer from the Demcon convergence Helpdesk?
Our Service & Support team aims to answer your question within one working day.
+
How can I contact the Demcon convergence Helpdesk?
The Demcon convergence Helpdesk is available from Monday to Friday from 8:30 – 17:00 (CE(S)T)
E: helpdesk@con-vergence.com
P: +31 (0)53 461 55 57
M:+31 (0)6 8296 6614 (also WhatsApp)
+
What specific information does the Demcon convergence Helpdesk need?
In order to provide you with the best Service & Support possible, we will need the following information:
1. Contact details
2. Serial number
3. Detailed problem description:
– Is it a software or hardware problem?
– In case of software, please explain your problem in detail.
– In case of hardware, which specific part is affected (e.g., strange sounds, leakage)?
4. If possible, please send us screenshots, pictures or video’s of the problem/parts.
5. If the question is program related, please include the program.
+
How does the TeamViewer program on my desktop work?
The helpdesk uses Teamviewer software to remotely assist you with your system. Lots of tutorials on this tool are available online
A session will be scheduled whenever remote assistance is necessary. Instructions will be send to you by e-mail. For a smooth session, please make sure that:
Every session is maximized to 30 minutes per session.
+
How do I setup a Service Level Agreement for my Demcon convergence Inspector?
+
What does the red light on Bronkhorst Flowmeters indicate?
The Bronkhorst flowmeters are equipped with lights that indicates their status. First, the red light can indicate an error. But it could also indicate communication activity. Whenever the software is running, communication is taking place. Hence, the red light will ligt up. This is normal behaviour.
Errors will be shown in the software under Program Information and Warnings.
+
What user maintenance is advised for my Demcon convergence Inspector?
Depending on fluids/gases or particles used during the experiments, the Demcon convergence Inspector may be subjected to fouling. To properly clean the fouling, the user can perform occasional flushes or cleaning cycles to negate the fouling effects. Below you will find a table of cleaning methods.
F0uling material | Active cleaner | Medium | Concentration | pH | Method |
---|---|---|---|---|---|
Biofouling: algea, bacteria | NaOCl | (demi)water/basic solution | 250-500 ppm | pH10 | Recirculate affected parts with solution, flush out with neutralizer/(demi)water |
Scaling: lime/mineral deposits | HCl | (demi)water | – | pH2 | Recirculate affected parts with solution, flush out with neutralizer/(demi)water |
Iron oxides/metallic oxides | Citric Acid | (demi)water | – | pH4 | Recirculate affected parts with solution, let sit for short/medium periods, flush with neutralizer/(demi)water |
A small batch of cleaning solution can be prepared and then recirculated throughout the system. Check the flowsheet, open all necessary valves and let the pump(s) run on a gentle speed. Make sure the outlet returns in the cleaning solution and let the system run for as long as necessary. When finished, flush out the cleaned areas to remove the cleaning solution again.
Some components or connectors rely on rubber or plastic O-rings. Some of them are heavily subjected to wear and tear. Check the (visible) O-rings periodically for wear and tear and replace them when necessary.
Some of the larger pumps have oil reservoirs to lubricate and cool moving parts. It is essential for these to have a sufficient oil level in the reservoir. Usually, these have a cap where the oil can be filled and the oil level can be checked. For more information about the type of oil or instructions, please contact the Demcon convergence Helpdesk or refer to the pump manual.
+
How do I store my Demcon convergence Inspector for longer periods of time?
Storing your Demcon convergence Inspector for longer periods of time is possible. Please make sure that the fluids are mostly out of the system. First, clean the system. Second, flush it with clean water and finally empty out the fluids from the system. In most cases, the pumps can empty out most fluid on their own when not connecting the input tubing with fluid. But if this is not deemed sufficient, pressurized air could be used to clean out the remains.
+
What is auto-zeroing for coriolis flowmeters?
Flowmeters that rely on the coriolis measurement principle, have what is called a zero stability error. When there is no flow, ideally, the meter should display around zero flow. The zero point can start to deviate over time when fouling has occurred. In that case the zero procedure must be carried out. The auto-zero procedure determines the resonance frequency of the measuring tube in the flowmeter and it will set the flow back to zero again.
The auto-zero procedure must be performed under the correct circumstances and must not be overused.
Aim for these conditions when auto-zeroing:
+
When should auto-zeroing for coriolis flowmeters be used?
Fouling, a change of medium or a change in temperature might affect the zero-point. If that is the case, please start the auto-zeroing procedure. This procedure should not be overused and might be of no need.
+
How can I remove a blockage from one of the components?
Because of the nature of testing certain DUT’s with fluids containing possible fouling components, it is not unlikely for a blockage to occur. Blockages can often be foung in components with small orifices. If a blockages is ascertained, try the following:
+
Where can I find the instruction manual?
An instruction manual will be provided with your system. A copy of the manual is available at the helpdesk.
+
Where can I find the software manual of my Demcon convergence Inspector?
The latest version of the software manual is included on the PC delivered with the Demcon convergence Inspector. It can be accessed by clicking the ‘User Manual’ button in the software or it can be found in C:\Program Files (x86)\Demcon convergence Inspector\Config.
A hardcopy of the software manual is included with the delivery of the Demcon convergence Inspector system. On request, the manual is available at the Demcon convergence Helpdesk.
+
I forgot my password, help!
When you use the Demcon convergence Inspector for the first time, the password is osmo. If this password was changed by you or your colleagues, there is nothing much we can do.
The guest user is always used for first time use and will have administrator right. It has no password.
In case a password for a specific user account got lost, it can be accesed by making a copy of the user account folder. Rename that folder to a new username and make a new username with the same name as the folder and a new password. Now you should be able to log in again!
+
What to do when I receive a verification connect error?
Inc ase of a verification connect error, there are two possibilities:
+
Where is the license file for my Demcon convergence Inspector installed?
The location of you license.lic file depends on the software version. You can find you license file in the next directories in your computer:
Software versions V6 and below:
/OSMO/Users/
Software versions V7 up to V7.x.x.x.:
/Inspector/Users/
Software V7.x.x.x. and up:
/Inspector/Users/
Once you received a new license, you need to overwrite the old license in the corresponding directory.
+
How do I write a program file for my Demcon convergence Inspector?
Please refer to QA-M0003v4 – software and programming manual.
+
Can I backup my system files?
The systems creates a folder under User/Documents/ with the name OSMO or under Public/Documents with the name Inspector by default. Every user will have access to this folder with several subfolders:
The user can also select his own default folder location for data, reports, programs and quick-programs in the software settings.
It might also be a good idea to back-up the software installation. The appointed folder can be found under:
C:\Program Files (x86)\Demcon convergence Inspector, or C:\Program Files (x86)\OSMO Inspector.
Do you have a question that is not listed in this section? Please contact our Helpdesk!
Dont miss out on our latest developments!
You will be the first to know about our newest products, exciting projects and more if you subscribe to our awesome newsletter.